GIMA SpA
6.1 GENERAL
No user adjustable parts are within the equipment, either for calibration or service purposes. The equipment housing must not be
opened: the warranty is invalidated by any unauthorized entry into the unit. In the event any repair or adjustment work being
necessary, the whole equipment should be returned to the LED S.p.A. Service Centre 04011 APRILIA (LT) - ITALY, or to a other
Authorized Centre, together with a description of the fault.
Maintenance work by the user is mainly the cleaning of the exterior of the cabinet, cleaning and sterilization of the accessory items
and checking of the equipment before each use. Carrying out function and safety check for verification of the parameters is
demanded to specialized technical people.
6.2 CLEANING OF THE CABINET
Switch the equipment off complitely and disconnect the mains supply before any cleaning is undertaken. Clean the outside of the
cabinet with a damp cloth. No chemical should be used; a mild non abrasive cleanser may be used when necessary.
6.3 CLEANING AND STERILIZATIONOF THE ACCESSORY ITEMS
The best thing to do is to use only one time use accessories and discard them after use. Since some of the accessory items are to be
used more than once it is mandatory to clean carefully and sterilize those accessories before the new use. The best way to clean and
sterilize the reusable items is to follow the direction of the supplier of each
item. When original reusable accessories supplied by LED S.p.A. are applied, the cleaning by using soft cleanser and sterilization
through steam sterilization at 134 °C is recommended.
6.4 CHECKING OF THE EQUIPMENT BEFORE EACH USE
Each time the use of the electrosurgical equipment is planned a check of the most important safety aspects has to be implemented
considering at least the following:
§
Check the integrity of cords, connections, wire breakage, etc.
§
Assure that all the electrical equipment is properly grounded
§
Assure that all the accessories that should be used are available and sterilized.
Manuale d'Istruzione / Instruction's Manual / Manual de Instrucciones
6. PREVENTIVE MAINTENANCE
MA126IGBa
DIATERMO 106
32