Technical Specifications - Wellell Optima Auto Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 30
Optima Auto Series / English

10. Technical Specifications

Item
Power Supply (Note: See
rating label on the product)
Fuse Rating
Cycle time
Dimension (L x W x H)
Weight
Atmospheric
Pressure
Temperature
Environment
Humidity
Classification
Communication
AC 220-240V 50 Hz, 0.6A
AC 110-120V 60 Hz, 1A
T1AL, 250V
10, 15, 20, 25 minutes
34.1 x 16.5 x 26.0 cm
5 Kg /11 lbs (with battery)
700 hPa to 1060 hPa
Operation: 10°C to 40°C (50°F to 104°F)
Storage: -15°C to 50°C (5°F to 122°F)
Shipping: -15°C to 70°C (5°F to 158°F)
Operation: 10% to 90% non-condensing
Storage: 10% to 90% non-condensing
Shipping: 10 % to 90% non-condensing
Class II, Type BF, IP21
Applied Part: Air Mattress
Not suitable for use in the presence of a flammable anesthetic
mixture (No AP or APG protection)
Cellular
*Twelve Bands FDD-LTE: 700, 800, 850, 900,
1700/2100 (AWS), 1800, 1900, 2100, 2600 MHz
(bands 1, 2, 3, 4, 5, 7, 8, 12, 18, 19, 20, 28)
* Seven Bands UMTS (WCDMA/FDD): 800, 850, 900,
1700/2100 (AWS), 1800, 1900 and 2100 MHz
(bands 1, 2, 4, 5, 8, 9, 19)
*Quad Band GSM: 850, 900, 1800 and 1900 MHz
Bd1, Bd2, Bd3, Bd4, Bd5, Bd7, Bd8, Bd12, Bd17, Bd18, Bd19, Bd20,
Bd28
RF ID No: FCC ID: QIPPLS62-W
Note: The device should be used at a minimum
distance of 8" (20 cm) from the body duration
operation.