Battery Life Information
Durable Period (Service Life)
Blood Pressure Measurement
Measuring Method
Measurement Range
Accuracy
Inflation
Deflation
Measurable Wrist Circumference
Maximum temperature of the applied part Lower than +43°C (Non-Charging Mode)
This device complies with IEC 80601-2-30, and has been clinically investigated according to the require-
ments of ISO 81060-2:2018/A1: 2020. In the clinical validation study, K5 was used on more than 85
subjects for determination of diastolic and systolic blood pressure.
Typical usage: 7 days
Typical battery life: A new battery on a full charge will last for
approximately 500 cycles when operating at room temperature
(25°C).
The battery level can be seen on the device screen.
Device body: 3 years; airbag & strap: 1 year. (Product Service
life is normally considered as the time from first use until the
device ceases to fulfil its intended use.)
Oscillometric method
Pressure: 0–300 mmHg (0–40 kPa)
Pulse rate: 40–180 times/min
Diastolic blood pressure: 40–130 mmHg (5.3–17.3 kPa)
Systolic blood pressure: 60–230 mmHg (8–30.7 kPa)
Pressure: ±3 mmHg (±0.4 kPa)
Pulse rate: ±5%
Automatic pressure pump
Automatic fast deflation
M-sized airbag: 13.0–16.0 cm/5.12–6.30 in
L-sized airbag: 16.1–20.0 cm/6.34–7.87 in
Please select the appropriate airbag according to your mea-
sured wrist circumference.