WARNING!
Risk of Falling!
Excessive or intermittent loads can lead to
spontaneous loss of suction force.
– Before using the suction grab bar, check the
suitability for the relevant user and his/her
usage behaviour.
– The maximum load that can be applied to the
suction grab bar is 30 kg.
2.2 Labels and Symbols on the Product
Identification Label (Fig. 2)
Conformity mark
Lot number
Manufacturer
Warning Label
Warning label for safe usage. See
hazard statements in chapter 4 Setup
(Fig. 4), page 3 .
3 Product Overview
3.1 Intended Use
Aquatec Stileo are grab handles in various lengths working
with two suction cups ensuring fixation on a surface by
vacuum force. There are two indicators providing information
about remaining strength of the suction cups ensuring the
support.
This product is intended to be used as a transfer aid e.g.: to
enter into or get out of a bathtub. It is designed to provide
support during transfer but not designed to hold a person's
entire body weight.
Intended User
Users are any adults and adolescents with restricted or
limited mobility or limited muscle strength. If used without
an assistant, the users must be able to keep their own
balance.
Indications
Restricted or limited mobility (e.g.: legs and/or hips) or
having balance issues to stand up or walk due to
transitional or permanent disabilities.
There are no contraindications known when using the
product as intended. For the maximum load, see 8 Technical
Data, page 4 .
3.2 Scope of Delivery
The scope of delivery is shown in figure 1.
4 Setup (Fig. 4)
The suction grab handle can be installed on sealed and
smooth surfaces.
1526478-G
1. Release both folding levers A.
2. Press and hold the suction grab handle evenly with both
3. Close both folding levers.
Maximum safe
4. To uninstall, remove the suction grab handle in reverse
load
Read user
manual
Medical Device
Observe warning label regarding the indicators:
5 Usage
1. Grip the handle with one hand and carry out the desired
6 Maintenance
6.1 General Maintenance Information
This product is maintenance-free, provided that the cleaning
and disinfection instructions are adhered to.
6.2 Cleaning and Disinfection
General Safety Information
WARNING!
Risk of Falling!
Incorrect mounting may lead to the suction grab
handle suddenly detaching from the surface. The
indicators on the suction grab handle allow the
user to check if it is mounted correctly (it is not a
vacuum indicator!).
– Before installation ensure that the suction grab
handle is undamaged, free of oil, dust and dirt.
suction pads B against a suitable surface.
order.
Lift the suction pads B for easier removal (e.g. with
a finger).
WARNING!
Risk of Falling!
There are two indicators C present on the suction
grab handle displaying green when the grab
handle is installed correctly and stable.
– If one or both indicators display red, the suction
grab handle is not safely installed and must be
repositioned.
Red hand: Do not use the product
when one or both indicators display red.
Green check: Only use the product
when both indicators display green.
WARNING!
– Before every use, check the product for damage
and ensure that it is securely mounted.
– Make sure the indicators both display green
before using the product.
body movement.
CAUTION!
Risk of Contamination
– Take precautions for yourself and use
appropriate protective equipment.