Descargar Imprimir esta página

Steinberg Systems SBS-DK-1000 Manual De Instrucciones página 43

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 135
c)
Connect the power supply cable to the power socket.
d)
Switch on the device with I/O button.
e)
Start mixing and set the appropriate speed following the "Mixing speed
regulation" chapter.
e)
Start heating and set the appropriate temperature following the
"Temperature setting" chapter.
f)
After completing operation of the device, switch off heating and mixing,
switch of the device with I/O button, and disconnect the device from the
power source.
Liquid temperature measurement
a) Remove the internal thermocouple from the socket located in rear portion of the
device.
Internal thermocouple
b) Connect the external thermocouple to the socket in the rear portion of the device.
The display will indicate the ambient temperature PV and set temperature SV.
c) Insert the metal part of thermocouple into the liquid for which temperature will be
measured. The display will indicate the measured temperature PV and set
temperature SV.
d) The safeguard will be activated when the external thermocouple is disconnected. In
such a case the display will show "hhhh" message, the green indicator light switches
of and the device stops working. When the above situation occurs, check the technical
condition of the device.
e) If it is not necessary to use external thermocouple, disconnect it and reconnect the
internal thermocouple.
Assembly of external thermocouple
Assembly of external thermocouple should be performed when it is necessary to heat
the liquid and its temperature must be monitored.
a)
Screw in stand into the opening in the corner of the device housing.
b)
Apply the cross connector on the stand and screw on the clamping knob.
c)
Install the boom with handle in in the cross connector and screw in the
clamping knob.
External thermocouple

Publicidad

loading

Este manual también es adecuado para:

Ex10030661