ELECTRICAL SPECIFICATIONS
Accuracy: ± (a% of reading + b digits).
Ambient temperature: 23°C ± 5°C (73.4°F ± 9°F) Relative humidity: ≤75%
Caution:
To ensure measurement accuracy, the operating temperature should be within 18°C~28°C and the
fluctuation range should be within ±1°C. When the temperature is <18°C or >28°C, add temperature
coefficient error: 0.1 × (specified accuracy)/°C.
1. DC voltage
Range
60.00 mV
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
•
Input impedance: About 1 GΩ for mV range, about l0 MΩ for other ranges
Accuracy guarantee: 1%~100% of range; short circuit allows least significant digit ≤5
•
•
Max input voltage: 1000 V (if the voltage is >1000V, the red indicator light will be on and the buzzer
will sound an alarm; if the voltage is >1010 V, the LCD will display "OL")
•
Overload protection: 1000 V
2. AC voltage
Range
60.00 mV
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
LoZ AC V 600.0 V
•
Input impedance: About l0 MΩ
•
Display: True RMS
•
Frequency response: 40 Hz ~ 1 kHz
•
The AC crest factor can be ≤3.0 at 3000 counts, and can only be ≤1.5 at 6000 counts. The additional
error should be added according to the crest factor for a non-sinusoidal wave, as follows:
a. Add 4% when crest factor is 1~2
b. Add 5% when crest factor is 2~2.5
c. Add 7% when crest factor is 2.5~3
16
button or insert the USB module.
USB
Resolution
0.01 mV
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Resolution
0.01 mV
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
0.1 V
Accuracy
± (0.8%+5)
± (0.8%+3)
± (0.5%+3)
± (0.5%+3)
± (1.0%+3)
Accuracy
± (1.2%+5)
± (1.2%+5)
± (1.0%+3)
± (1.0%+3)
± (1.0%+3)
± (1.2%+5)
± (2.0%+5)