Frequency
Frequency
Operation and
maintenance manual
Failure to comply with the above instructions could cause:
detaching of paint with possible accidental dropping of such paint into
-
the patient area;
early ageing of the plastic parts with consequent weakening and the
-
possibility of breakages;
tarnishing of the protection screens and glass.
-
6.2 Cleaning the Product
We recommend you to clean the Product every day.
Possibility of damaging Product
Do not use sharp, pointed or abrasive objects, to avoid the risk of
-
damaging surfaces.
Do not pour liquids directly on the Product.
-
Clean the Product with a damp, but not wet, cloth.
-
Clean with suitable detergents with low alkaline content and
-
chlorine free. Do not use abrasive products, petrol, paint thinners,
alkaline detergents, acids, containing alcohol or aldehydes;
Dose the detergents strictly according to the percentage indications
-
shown on the manufacturer's technical sheet, being careful that no
liquids penetrate into the joints of the various Product parts, with
special care give to the reflector and supporting structure.
6.3 Product disinfecting
We recommend you to disinfect the Product every time before use.
Possibility of damaging Product
Disinfectants can contain substances that are harmful for the health; use
disinfectants indicated by the national commission for hygiene and
disinfection, according to the hygienic standards adopted by the
Responsible Organization.
- Do not use sharp, pointed or abrasive objects, to avoid the risk of
damaging surfaces.
MU_029_EN
07/11/18
Rev.0