See drawing 198
Operation and
maintenance manual
The handpieces can withstand about 200 steam sterilization cycles in
accordance with the following parameters:
steam sterilization at 121°C 1.3 bar for 25 to 30 minutes
-
steam sterilization at 134°C 2.3 bar for 4 minutes
-
Do not exceed a sterilization temperature of 134°C.
Strictly keep to the ISO 17665-1 standard
When placing in the autoclave, make sure the open side of the
handpieces is turned downwards. The handpieces must be free and not
burdened by other material being sterilized.
Damaged handpieces must no longer be used.
7 Adjustment and maintenance
7.1 Installation of swinging arm
The Product is sold already balanced and does not require further
adjustment. In the event of the spring swinging arm becoming stiff or
loose over time, mechanical intervention is possible by regulating the
compression of the internal spring.
Manually move the cover (1) forwards. Fit the adjustment pin (2) in the
holes of the ring nut and turn in the direction indicated by the arrows to
increase/decrease the load on the spring.
If the swinging arm drops, this means the elastic force of the spring is
insufficient:
- turn the lever upwards and release the spring.
If the swinging arm lifts up, this means the elastic force of the spring is
too high:
- turn the lever upwards and release the spring.
At the end of adjustment, manually reposition the cover (1) in its original
position.
The swinging movement of the arm can also be adjusted upwards.
The Product is sold with maximum set swinging movement. If the
swinging movement is to be reduced upwards, manually move the cover
(1) forwards and insert the adjustment pin (2) in the second ring nut. By
turning the pin downwards, the swinging movement can be reduced until
it is in horizontal position.
The swinging movement downwards cannot instead be changed.
MU_029_EN
07/11/18
Rev.0
Page 19 of 35