Blanking-perioden
Parameters
Waardebereik
Far-field bescherming na V
100 ... (10) ... 220 ms
s
Far-field bescherming na V
100 ... (10) ... 220 ms
p
Ventriculaire blanking na A
30 ... (5) ... 70 ms
p
PMT-bescherming
Parameters
Waardebereik
PMT-detectie/-beëindiging
UIT; AAN
VA-criterium
250 ... (10) ... 500 ms
Timing: 1-kamer
Basisfrequentie dag/nacht
Parameters
Waardebereik
Basisfrequentie
30 ... (1) ... 88 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 bpm
Nachtfrequentie
UIT
30 ... (1) ... 88 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 bpm
Nachtbegin
00:00 ... (10 min) ... 23:50 hh:mm
Nachteinde
00:00 ... (10 min) ... 23:50 hh:mm
Frequentiehysterese
Parameters
Frequentiehysterese
Repetitieve hysterese
Zoekhysterese
Standaard
06:00 hh:mm
Waardebereik
Standaard
UIT
UIT
-5 ... (-5) ... -90 bpm
UIT; 1 ... (1) ... 15
UIT
UIT; 1 ... (1) ... 15
UIT
Bovengrensfrequentie
Standaard
Parameters
100 ms
Bovengrensfrequentie in de
modus VVT(R)
150 ms
30 ms
Refractaire periode
Parameters
Refractaire periode
Standaard
AAN
Stimulatie en sensing: 2-kamer
350 ms
Impulsamplitude en impulsduur
Parameters
Impulsamplitude A
Impulsduur A
60 bpm
Impulsamplitude V
Impulsduur V
UIT
Gevoeligheid
Parameters
22:00 hh:mm
Gevoeligheid A
Gevoeligheid V
Atriale amplitudemanagement
Parameters
Atriale amplitudema-
nagement
Min. amplitude
Start stimulatiedrempeltest
192
Waardebereik
90 ... (10) ... 200 bpm
Waardebereik
200 ... (25) ... 500 ms
Waardebereik
0,2 ... (0,1) ... 6,0 ... (0,5) ... 7,5 V
0,1 ...(0,1) ... 0,5; 0,75; 1,0; 1,25;
1,5 ms
0,2 ... (0,1) ... 6,0 ... (0,3) ... 7,5 V
0,1 ... (0,1) 0,5; 0,75; 1,0; 1,25; 1,5 ms 0,4 ms
Waardebereik
AUTO
0,1 ... (0,1) ... 1,5 ... (0,5) ... 7,5 mV
AUTO
0,5 ... (0,5) ... 7,5 mV
Waardebereik
ATM (alleen monitoring)
AAN; UIT
0,5 ... (0,1) ... 4,8 V
2,4; 3,0; 3,6; 4,2; 4,8 V
Standaard
130 bpm
Standaard
250 ms
Standaard
3,0 V
0,4 ms
3,0 V
Standaard
AUTO
AUTO
Standaard
AAN
1,0 V
3,0 V