Power Supply And Battery Operation; Backup Battery; Charge Battery - Medela Thopaz+ Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 303

2 Power supply and battery operation

WARNINGS
This product contains lithium-ion batteries which bear risk of fire, explosion and burns.
Do not disassemble, crash, heat above 100 °C (212 °F), incinerate or dispose of in fire.
Thopaz + can be operated from the mains power supply
079.0037 or with the integrated lithium-ion battery. Before using the docking station, read
first the docking station instructions
The battery duration is dependent on the run-time of Thopaz + . This is influenced by the extent of
parenchymal leakage and the set pressure. Thopaz + does not run continuously but only switches
on when the actual and nominal values differ.
During continuous operation, Medela guarantees a minimum run-time of 4 hours in battery
operation after the device has been fully charged. In practical operation the real battery run time
results in >10 hours.
If Thopaz + is not used very often, the battery must be charged approximately every 6 months
to ensure optimum functioning.

Backup battery

If an internal defect occurs (broken cable, battery defective) Thopaz + turns off and the
acoustic warning is sounded for at least 3 minutes (powered by a backup battery). Under
these conditions Thopaz + functions as a one way valve. Replace Thopaz + immediately.

Charge battery

200.1554. The battery is charged during mains operation.
Plug the Thopaz + power supply into a wall outlet and open the
small protection cover to access the DC socket. Plug the coaxial
power supply into the DC socket. Moving bars on the battery
display will indicate charging is occurring [
087.0059, from the docking station
].
Power supply and battery operation
| 11

Publicidad

Tabla de contenido
loading

Tabla de contenido