7.6

Check therapy progress

7.6.1

Main display

The current size of the air leak (ml/min) is shown on the left side.
The total amount of collected fluid (ml) during the therapy is
shown on the right.
According to your preference the air leak history of
12 h / 100 ml / min and two customized fluid parameters can be
displayed in addition. (To change factory settings please refer
to chapter 10 – Changing settings).
– Air leaks between 0 and 1000 ml / min are displayed in 10 ml
steps.
– Air leaks over 1000 ml / min are displayed in 100 ml steps.
– Rounding function
The digital display of air leak values is rounded to zero
ml / min (0 ml / min) if the air leak measurement indicates an air
leak < 5.0 ml / min. If the air leak measurement indicates an air
leak ≥ 5.0 ml / min and < 15 ml / min the digital display of air
leak values is rounded to ten ml / min (10 ml / min)
– The fluid measurements are displayed as follow: In a 0.3 l
canister in 1 ml steps, in a 0.8 l canister in 5 ml steps and in a
2.0 l in 10 ml steps. After a total drainage amount of 5000 ml
and more the display will switch to liter (l). Thopaz + will display
fluid values in 0.01 l steps (equals 10 ml steps) thereupon.