LipoSurg
6
Starting up
6.1
Device setup
● Installation layout
● Place the LipoSurg and all required accessories and instruments on an even, non-slip surface and make
sure you have good access to all controls.
● The installation of the device in close proximity to other devices is prohibited due to EMC – please see
section 3.1 and the manufacturer's EMC declaration in the appendix of this manual.
● Do not allow the operating range of the device, including cable, motor and Conform Cannula Handpiece,
to be compromised by limiting factors.
● The systems display must be fully visible at all times.
● The pedal must be placed within stepping distance between the patient and the surgeon.
● It must be explicitly ensured that no objects can fall on the pedal.
● The power plug at the rear of the device must be accessible at all times.
● The motor ventilation slots must be kept clear in order to prevent the motor temperature from becom-
ing excessive. Ventilation channels in the handlebar prevent the electronic motor from overheating.
6.2
Connection to the power supply
Before plugging the power cable into the power socket for the first time, you must check the supply volt-
age setting next to the power switch.
If the voltage shown does not correspond to the local mains voltage, the grey fuse holder must be set to the
correct voltage:
A) Unplug the power cable.
B) Use a screwdriver to open the fuse slot.
C) Remove the fuse holder.
D) Remove the grey fuse holder and reinsert it so that the local mains voltage setting is shown in the small
window.
E) Slide fuse holder back in and close the fuse slot.
F) Check the mains voltage shown on the fuse slot.
G) Plug the power cable back into the device.
In order to prevent the risk of electric shock, the device may only be connected to a power network with
a PE protective ground conductor.
Only a tested power cord may be used to connect the device to the power supply.
Operating instructions LipoSurg, REF 31648, V48/18
EN
9