LipoSurg
7.7
Functional check
Prior to every startup of the LipoSurg, or the use of accessory equipment, the user must always ensure that
each individual component is in good working order, free from defects, and is clean, sterile and operational.
All inscriptions on the device and its accessories must be readable and there must be no loose parts in the
device. Once the device is switched on, the most recent settings entered appear on the display.
7.7.1
Electronic motor
The functional check of the electronic motor is carried out without the Conform Cannula Handpiece
attached. To activate the liposuction mode press the key "LIPO". Press the key "LIPO " to increase the
speed to its maximum value. Press the pedal. The electronic motor starts running and accelerates up to its
maximum speed. When releasing the pedal, the electronic motor decelerates and finally comes to a stop.
The motor ventilation slots must be kept clear in order to prevent the motor temperature from becoming
excessive. When the electronic motor is placed in the handle, built in channels ensure sufficient cooling
of the motor.
7.7.2
Peristaltic pump
Activate the infiltration mode by pressing the "INFIL" key. Press "INFIL" to select the maximum speed
of the pump (100 %). Press the pedal all the way down. The pump starts running and quickly speeds up to its
maximum speed. At the outlet of the tube respectively at the infiltration needle or at the infiltration cannu-
la the liquid is spurting out.