7. Operating Instructions
1. Leg adjustment
The legs on the Hoyer Presence are electrically adjustable for width. The legs can be opened to
enable access around armchairs or wheelchairs. For transferring and negotiating narrow doorways
and passages the lift legs should be in the closed position.
Electric leg adjustment - is achieved by pressing the appropriate buttons on the hand control.
The legs will be locked whenever the hand control switch is released.
2. Castors and Braking
The lift has two braked castors which can be applied for parking. When lifting, the castors should
be left free and un-braked. The lift will then be able to move to the centre of gravity of the lift. If
the brakes are applied it is the patient that will swing to the centre of gravity and this may prove
disconcerting and uncomfortable.
3. Raising and lowering the boom
The movement of the boom is achieved by a powerful electric actuator, which is controlled by
a simple hand control unit. The hand control has two buttons with directional arrows UP and
DOWN. The actuator stops automatically at the limit of travel in both directions. The hand control
plugs into a socket at the base of the control box. There is a hook on the rear of the hand control,
which allows it to be "parked" on the mast or boom when not in use.
4. Emergency Stop
The red Emergency Stop Button is located on the front of the control box and is activated by
pressing in. This will cut all power to the lift and only be reset by twisting the button counter
clockwise and releasing.
5. Redundant controls
All Hoyer lifts are fitted with raise and lower buttons on the control box. These are located
underneath the emergency stop button and can be used to lower/raise the patient should the
hand control fail.
6. Mechanical Emergency Down
In the case of a complete electrical failure the electrical actuator is fitted with a mechanical
lowering device (RED BOSS). Dependent on your lift, the device is either pulled upwards, or
twisted by hand to activate. A slow decent will commence. Repeat this process until the patient
has been safely lowered.
If this feature is used, the lift MUST be subsequently checked out by a competent engineer.
7. Batteries
The batteries are protected from deep discharge by a LOW VOLTAGE ALARM. This will
sound when the batteries need recharging and the hand control is being operated. It will not
sound independently of the hand control being operated. CAUTION: DO NOT IGNORE THIS
WARNING ALARM. Complete the lifting operation and place the lift on charge (see charging
instructions).
To avoid possible permanent damage to the battery, the battery should be placed on
charge as soon as the display indicates the half empty battery symbol.