Data Security
and Encryption
options (at rest)
Data Security
During Repair
Saving files to
the Network
Log Out
Software
Updates
P23566L
Disable the "Patient Data Export" option when transferring archived cases
from the device to external media. Go to the Save Settings screen and
uncheck "External Media Options. (see the Save Settings section).
If "Patient Data Export" capability is activated, enable the "Encryption"
option when transferring and storing archived data to external media. (see
the Save Settings section).
When returning the device to Stryker for repair, remove all PHI, per Stryker
policy. See contract for terms of agreement. Contact a Stryker sales
representative to request a Stryker factory restore.
If transferring the device with archived PHI, anonymize the data before
transit. Go to Settings >System > Data Protection (see the System Data
Protection Settings).
Always select secure transmission of data to the network. (see the Saving to
Network section).
Always log out manually from the device once the procedure is complete.
Warning: the device does not contain an automatic log off feature.
Install software updates provided by Stryker as they are released.