Accuracy
Refer to the "Specifications" section for operating conditions and temperature
coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed
amount) at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 75% relative humidity.
Accuracy is specified from 5% to 100% of the range unless noted otherwise.
AC Voltage
Range
(40 Hz to 400 Hz)
200.0 V
600 V*
* 600 V range is specified from 20% to 100% of range.
DC Voltage
Range
20.00 V
200.0 V
600 V*
* 600 V range is specified from 20% to 100% of range.
Resistance
Range
200.0 Ω
2000 Ω
20.00 kΩ
200.0 kΩ
2000 kΩ
12
Accuracy
± (1.2% + 1.0 V)
± (1.2% + 10 V)
Accuracy
± (0.8% + 0.05 V)
± (0.8% + 0.5 V)
± (1.0% + 5 V)
Accuracy
± (1.2% + 0.5 Ω)
± (1.0% + 5 Ω)
± (1.0% + 0.05 kΩ)
± (1.0% + 0.5 kΩ)
± (1.2% + 5 kΩ)
Input Impedance
450 kΩ
Input Impedance
1 MΩ
Open Circuit Voltage
< 3 VDC