08.1 Û ELECTROMAGNETIC COMPATIBILITY EN 60601-1-2
WARNING: Interference with
other equipment. Though compliant
with the standard IEC 60601-1-2, the
PIEZOSURGERY® touch may nonetheless
interfere with other devices nearby.
The PIEZOSURGERY® touch must not
be used near to or stacked on other
devices. However, if this were to prove
necessary, you must check and monitor
correct operation of the device in that
configuration.
WARNING: Portable and mobile radio
communication appliances may affect the
correct functioning of the device.
Guidance and manufacturer's declaration - Electromagnetic emissions
The PIEZOSURGERY® touch is intended for use in the electromagnetic environment specified
below. The customer or user of the PIEZOSURGERY® touch should ensure that it is used in
such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
equipment. An electrical scalpel or
other electro-surgical units near the
PIEZOSURGERY® touch device may interfere
with its correct operation.
EMC precautions and must be installed
and activated in accordance with the EMC
information given in this paragraph.
Compliance
Electromagnetic environment – Guidance
Group 1
The PIEZOSURGERY® touch only uses RF energy
for internal function. Therefore, its FR emissions
are very low and are not likely to cause any
interference with nearby electronic
equipment.
Class B
The PIEZOSURGERY® touch is suitable for
use in all establishments, including domestic
establishments and those directly connected to
Class A
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Complies
08 Û TECHNICAL DATA
WARNING: Interference from other
WARNING: The device requires specific
EN