Descargar Imprimir esta página

Chauvin Arnoux AEMC 6503 Manual De Instrucciones página 14

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

English
If this instrument is damaged or a part is missing, please contact the seller
immediately.
This instrument is protected against accidental voltages up to 600V with respect
to earth in measurement category II, 300V in category III.
Failure to observe the instructions or the precautions for use may compromise the
protection provided by the instrument.
Please refer to this manual for each danger symbol encountered.
To avoid an electric discharge, injury, or damage to this instrument, and make
sure that you are using the megohmmeter in a risk-free manner, follow the safety
recommendations listed below:
This instrument can be used indoors or out (IP52) in environments where
the degree of pollution does not exceed 2, at an altitude of less than
2000m, with voltages up to 600V (CAT II) or 300V (CAT III).
Never use on a network of which the voltage or overvoltage category
exceeds the stated values
Before each use, check the integrity of the housing and of the insulation
of the leads and accessories. Replace any damaged cords. Do not use
the instrument if it seems damaged.
Use leads and accessories satisfying safety standards (IEC 61010-031)
for voltages and overvoltage categories at least equal to those
mentioned.
The use of suitable personal protective equipment is strongly
recommended. Avoid working alone.
Resistance and insulation measurements must be made only on circuits
that are isolated and not live.
Following an insulation measurement, leave the measurement leads
connected for a few seconds before disconnecting in order to discharge
the circuit tested.
Disconnect the cords from the instrument before cleaning, changing the
fuse, or opening the housing.
14
PRECAUTIONS FOR USE
C.A 6503

Publicidad

loading

Este manual también es adecuado para:

C.a 6503