TO ADJUST THE ACTIVE THERAPY PARAMETERS

1 .
Select the "Rotation" button .
2 .
Adjust the therapy parameters as described in step 2
of "To Start Rotation Therapy" .
Note:
• The adjusted parameters take effect as soon as
they are changed .
• If the angle of the head elevation is between 30 and
60 degrees, the maximum rotation settings, other
than the Turn Assist, are automatically reduced to
ensure safe rotation levels . If Rotation Therapy has
started and the head angle is raised to 60 degrees
or greater, Rotation will stop and the message
"Rotation Stopped Head Elevation Too High" will be
displayed on the screen .
• If rotation is stopped for any reason, "Rotation"
and "Stopped" are briefly displayed and the patient
returns to the supine position .
• If Rotation is active and "Max inflate" or "Turn Assist"
button is selected, the screen will display "Rotation
Paused" then "Maximum Inflate" or the Turn Assist
Screen will display .
• The head of bed (HOB) elevation angle is determined
by the sensor in the InTouch bed . If the support
surface and InTouch sensors do not coincide by
at least 10 degrees, the HOB elevation angle is
determined by the sensor in the support surface .
Always secure the support surface to the bed frame
to ensure the sensors are aligned .

TO STOP ROTATION THERAPY

1 .
Select the "Stop" button . The patient returns to the
center position and the rotation stops .