Recommended separation distances between portable and mobile RF
communications equipment and the VenaFlow Elite
The VenaFlow Elite is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled . The customer or the user of the VenaFlow Elite can
help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the VenaFlow Elite
as recommended below, according to the maximum output power of the communications
equipment .
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer .
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies .
NOTE 2: These guidelines may not apply in all situations . Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people .
26
Separation distance according to frequency of transmitter
150 KHz to 80 MHz
d =
0,12
0,38
1,2
3,8
12
m
80 MHz to 800
MHz
d =
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d =
0,23
0,73
2,3
7,3
23
ENGLISH