Illustrations in this user manual are for reference only. Please refer to the actual product.
The silicone anti-return valve, silicone massage cushion and tube are wear parts. If you notice wear and tear, purchase replacement parts from an authorised
retailer.
Wash the non-return valve very gently and do not allow food to dry inside it.
NOTE! Always remove the anti-return valve by carefully pushing its rim with your finger through the diaphragm cup. Never grasp the tip of the valve, as it can
be damaged what can weaken the suction power.
1.
Nipple correction mode / nasal aspirator mode
7.
2.
Level up
8.
3.
Vacuum level indicator
9.
4.
Level down
5.
Suction frequency indicator (high, medium, low)
10.
6.
Mode button (stimulation, expression, bionic)
11.
1. ON/OFF button
: Press and hold for approximately 2 seconds to turn the device on or off.
2. Mode change button:
Press the button once to change the mode.
You can choose one of the following modes:
• Stimulation mode stimulates milk let-down and has 9 levels of suction. After running for 2 minutes in the stimulation mode, the breast pump will automatically
switch to the expression mode at the first level of suction. You can change a mode to stimulation or expression at any time by pressing the button.
• The expression mode is designed specifically for milk expression. Also in the expression mode, you can choose one of 9 suction levels.
• Bionic mode imitates baby's sucking. It is a combination of stimulation and expression modes that work alternately, and it has 9 levels of strength.
3. Pause
: Press the pause to stop the device, and press again to resume. The device turns off automatically after 3 minutes from pressing the pause.
4. Suction strength: You can press the "^" to increase the suction power or "v" to lower the suction power. You can adjust the suction power for each breast
separately.
5. Frequency button
: There are three levels of suction frequency available (low, medium and high frequency). The default value is medium frequency.
6. Single/double mode: Press the button
to switch between single breast and double breast modes. If you want to express milk from both breasts,
connect the tube and accessory kit to the port marked with the icon
7. Flat nipple correction mode: Effectively helps to stimulate concave nipples. To activate the flat nipple correction mode press the button
you can adjust the suction level from 1 to 3.
NOTE! Take special care when assembling and disassembling the breast pump to avoid damage to its components.
1. Place the silicone massage cushion (fig. 1a) in the funnel. Stretch its edges over the edges of the funnel (fig. 1b) so that they are fitted tightly.
2. Place the diaphragm (fig. 1c) in the cup at the top of the body (fig. 1b) so that they fit precisely. Press the edges of the diaphragm. Then place the diaphragm
cap (fig. 1d).
3. Place the valve (fig. 1e) in the hole on the underside of the breast pump body (fig. 1b) by pushing it to the end. Screw the adapter onto the bottle (fig. 1f).
Screw the whole set to the underside of the body (fig.1b) so that they form one unit.
4. Connect one end of the tube (fig. 1h+i) to the top cover (fig. 1d) and the other end (fig. 1h+i) to the pump (fig. 1j). Make sure the tube is not tangled.
5. Place the funnel cover (fig 1g). The device is ready.
NOTE! Before first use, fully charge the battery. The breast pump has a rechargeable lithium-ion battery and does not require AA batteries or a permanent
connection to the mains. Once fully charged, stop charging to avoid damaging the battery. To extend the life of the battery, do not let it run down fully.
The battery is charged via the power adapter. To charge the battery, connect one side of the charging cable to the breast pump and the other part to the power
adapter. Then plug the power adapter into a power socket. It takes about 180 minutes to fully charge the breast pump's battery. When the battery is fully
charged, the battery LED indicator will stop flashing. The breast pump can be charged at any time. You don't have to wait for the battery to be fully discharged.
On a full charge, the breast pump can run for about 180 minutes.
The breast pump has a built-in timer and automatically switches off after 30 minutes of operation. If you want to restart the device, press the ON/OFF button.
Before each use:
• Wash your hands thoroughly.
• Make sure that the breast pump was pre-cleaned and dried. If not, follow the section Cleaning and Sterilising.
• Make sure that there is no dirt, liquid or moisture in the silicone tubing. Never turn on the breast pump if you notice water in the silicone tube.
Warning! Do not mouth blow the silicone tube to remove water.
Place the breast pump on your breast so that the nipple is deep inside the funnel (exactly in the middle).
NOTE! The silicone cushion must cling tightly to the breast. Otherwise, the suction may be reduced.
1. Sit comfortably.
2. Make sure the batteries in the breast pump are charged.
3. Press the ON/OFF icon to start pumping. The breast pump will turn on to the 1 suction level in the stimulation mode. Then, after running for 2 minutes, it
will automatically switch to the expression mode (first level). The breast pump is equipped with a timer to keep track of the length of the pumping session. The
device automatically turns off after 30 minutes of use.
4. You can choose one of the available modes. After running for 2 minutes in the stimulation mode, the breast pump will automatically switch to the expression
mode at the first level of suction.
5. After using, disassemble the breast pump. Clean the parts according to the Cleaning and Sterilising section.
NOTE! The breast pump, silicone tubes with connectors, power adapter and USB cable cannot be washed or sterilised in water. To clean, wipe with a cloth.
NOTE! Before putting the breast pump away from the breast, make sure it is turned off. If necessary, you can put the breast pump away during pumping by
placing your finger between the pad and the breast (just as you do it when putting the child away).
12_212reb_instrukcja_laktator_elektr_double_ms001.indd 4-5
12_212reb_instrukcja_laktator_elektr_double_ms001.indd 4-5
CONTROL PANEL Fig. 2
Single or dual mode button
12.
Level down
ON/OFF button
13.
Battery indicator
Mode indicator (stimulation, expression,
14.
Operating time indicator
bionic)
15.
Breast pump / aspirator mode
Level up
16.
Frequency button
Vacuum level indicator
17.
Pause
on the breast pump base housing.
. In this mode,
ASSEMBLY Fig. 3
USER MANUAL
4
CLEANING AND STERILISING
Disassemble the breast pump. Before first and after each use, wash all parts that come in direct contact with milk, diaphragm cover (fig 1d), diaphragm (fig. 1c),
breast pump body with funnel (fig. 1b), silicone massage cushion (fig. 1a), funnel cover (fig. 1g), silicone anti-return valve (fig. 1e), bottle adaptor (fig. 1o) and
bottle (fig. 1f), in warm water with mild detergent and place them in boiling water for 5 minutes to ensure hygiene. Dry all components thoroughly before use.
Note! Do not leave components on the bottom of the pot. Stir the water during sterilisation to prevent damage to components.
The silicone components in contact with the milk, i.e. the diaphragm (fig. 1c), silicone massage cushion (fig. 1a) and silicone anti-return valve (fig. 1e), can be
sterilised in an electric steriliser.
How to disinfect the teat?
Before first and each use wash all parts of the product and disinfect. Place for 5 min in boiling water. This is to ensure hygiene. Clean before each use. Boiling and
hard water may create lime scale effect. Wash all parts thoroughly with clean water, including areas that may be invisible but where food has passed, to remove
any food residues. After using, remove the remaining food from the bottle and the teat, rinse, wash and disinfect in a steam/electric sterilizer or in boiling water.
Bottle can be washed in dishwasher (max. temp. 60°C). It is not recommended to place the teat in a dishwasher. Hard chemicals / detergents may damage the
product. Do not use any antibacterial agents. Inspect the feeding teat before each use and pull the feeding teat in all directions. Especially when the baby has
teeth. Throw away at the first signs of damage or weakness. Do not tighten up the bottle locking ring, as it may disturb the teat's ventilation. Always check for
leakage. Never enlarge the teat flow holes by any means, as it can damage the teat. Take particular care when heating in a microwave. Always stir heated food
to ensure even distribution of temperature and check liquid temperature before giving it to the child. When warming food in a microwave, remember to leave
the bottle open. Do not place product in hot oven or heat in a gas/electric cooker. Do not leave a feeding teat in direct sunlight or heat, or leave in disinfectant
("sterilising solution") for longer than recommended, as this may weaken the teat. It is recommended to replace a teat max. after 2 months of use. Retain the
instruction for later use. Bottle made from Polypropylene. Batch code is placed on the packaging.
For your child's safety and health
WARNING!
Continuous and prolonged sucking of fluids will cause tooth decay.
Always check food temperature before feeding.
Throw away at the first signs of damage or weakness.
Keep components not in use out of the reach of children.
Never attach to cords, ribbons, laces or loose parts of clothing. The child can be strangled.
Never use feeding teats as a soother.
Always use this product with adult supervision.
WARNING!
• Do not use this product for purposes other than those listed in this user manual.
• Do not place the product near sources of fire.
• Before using the breast pump, make sure that all parts were assembled correctly.
• If you feel breast or nipple discomfort, stop using the breast pump immediately.
• Wash your hands before using the product.
• Keep the product out of the reach of newborns and children.
• Do not drop or hit the product to avoid damage.
• Do not touch product elements with bare hands immediately after sterilisation in boiling water to avoid burns.
Do not resell a used device.
MILK STORAGE
1. Due to the active biochemical and microbiological processes occurring in the milk, it is important to maintain a constant temperature of storage.
2. Expressed milk store in bottles, special sterile bags or sealed containers with a lid (which should be disinfected before use).
3. Mark each container with the date.
4. Freeze only fresh milk. Do not fill a container fully as frozen breast milk increases its volume.
5. Breast milk can be heated in the same way as any other food (but never in the microwave as this may destroy valuable nutrients and antibodies).
6. Frozen breast milk can be defrosted overnight in the refrigerator and then warmed gradually in a bottle warmer or hot water.
7. After defrosting, shake the container or stir the milk to mix the separated fat particles with the remaining volume of milk.
8. DO NOT store milk on shelves placed on a refrigerator door.
9. NEVER refreeze previously defrosted milk.
10. DO NOT add fresh breast milk to defrosted milk.
11. DO NOT add fresh breast milk to frozen milk.
12. NEVER heat breast milk by boiling or putting it in the microwave.
13. ALWAYS check the liquid temperature before feeding.
14. Dispose of any milk that has not been used.
Type of breast milk
Temperature (°C)
+4°C
from +4°C to +15°C
Fresh milk
+19°C to +25°C (room temperature)
from +25°C to +37°C
Frozen milk
-20°C
Thawed milk
+4°C
from +25°C to +37°C
Recommended time of storage
up to 96 hours
maximum 24 hours
optimally up to 4 hours
maximum up to 4 hours
optimally 3-6 months
up to 24 hours (since total thaw)
up to 4 hours