The waveform is output using the present carrier and modulating waveform settings.
SCPI Command
[SOURce[1|2]:]AM:STATe{ON|1|OFF|0}
[SOURce[1|2]:]FM:STATe {ON|1|OFF|0}
Carrier Waveform Shape
– AM or FM carrier shape: Sine (default), Square, Ramp, Pulse, Triangle, Noise (AM only), PRBS, or Arbitrary
waveform. You cannot use DC as the carrier waveform.
– For FM, the carrier frequency must always be greater than or equal to the frequency deviation. Attempting to set a
deviation greater than the carrier frequency will cause the instrument to set the deviation equal to the carrier
frequency.
– The carrier frequency plus the deviation cannot exceed the selected function's maximum frequency plus 100 kHz. If
you attempt to set the deviation to an invalid value, the instrument adjusts it to the maximum value allowed with the
present carrier frequency. The remote interface also generates a "Data out of range" error.
Front Panel Operations
Press [Waveform]. Then select a waveform shape.
SCPI Command
[SOURce[1|2]:]FUNCtion <function>
The APPLy command configures a waveform with one command.
Carrier Frequency
The maximum carrier frequency varies by function, model, and output voltage, as shown here. The default is 1 kHz
for all functions other than arbitrary waveforms. Arbitrary waveform "frequency" is also set using the
FUNCtion:ARBitrary:SRATe command.
Front Panel Operations
Press [Parameter] > Frequency. Use the numeric keypad or the knob and arrows to set a desired value. If you use the
keypad, select a unit prefix to finish.
Keysight EDU33210 Series User's Guide
101