SCPI Command
[SOURce[1|2]:]FREQuency:CENTer {<frequency>|MINimum|MAXimum|DEFault}
[SOURce[1|2]:]FREQuency:SPAN {<frequency>|MINimum|MAXimum|DEFault}
Sweep Mode
You can sweep with linear or logarithmic spacing, or with a list of sweep frequencies. For a linear sweep, the
instrument varies the output frequency linearly during the sweep. A logarithmic sweep varies the output frequency
logarithmically.
The selected mode does not affect the sweep return (from stop to start, if one is set).
Front Panel Operations
Press [Sweep] > Type Linear.
SCPI Command
[SOURce[1|2]:]SWEep:SPACing {LINear|LOGarithmic}
Sweep Time
Sweep time specifies the number of seconds required to sweep from the start frequency to the stop frequency. The
instrument calculates the number of points in the sweep based on the sweep time.
Sweep time: 1 ms to 250,000 seconds, default 1 s. For a linear sweep in immediate trigger mode, the maximum total
sweep time (including hold time and return time) is 8,000 s. The maximum total sweep time for linear sweeps using
other trigger modes is 250,000 s; the maximum total sweep time for logarithmic sweeps is 500 s.
126
Keysight EDU33210 Series User's Guide