4. Select Marker Freq. Use the numeric keypad or the knob and arrows to set a desired value. If you use the keypad,
select a prefix unit to finish.
SCPI Command
[SOURce[1|2]:]MARKer:FREQuency {<frequency>|MINimum|MAXimum|DEFault}
Sweep Trigger Source
In sweep mode, the instrument outputs a single sweep when a trigger signal is received. After one sweep from the
start frequency to the stop frequency, the instrument waits for the next trigger while outputting the start frequency.
– Sweep trigger source: Immediate (default), External , Time, or Manual.
– With the Immediate (internal) source, the instrument outputs a continuous sweep at a rate determined by the total
of the hold time, sweep time and return time. The sweep time for this source is limited to 8000 seconds.
– With the External source, the instrument accepts a hardware trigger on the front panel Ext Trig connector and
initiates one sweep each time Ext Trig receives a TTL pulse with the specified polarity.
– The trigger period must be greater than or equal to the specified sweep time.
– With the Manual source, the instrument outputs one sweep each time the front panel [Trigger] key is pressed.
Front Panel Operations
Press [Trigger] > Source.
Keysight EDU33210 Series User's Guide