Front Panel Operations
To specify whether the instrument triggers on the rising or falling edge on the Sync Out connector, press [Trigger] >
Trig Out Setup. Then select the desired edge by pressing Trig Out.
SCPI Command
OUTPut:TRIGger:SLOPe {POSitive|NEGative}
OUTPut:TRIGger {ON|1|OFF|0}
Frequency List
In frequency list mode, the instrument "steps" through a list of frequencies, dwelling on each frequency for a
specified period. You may also control progress through the list with triggering.
– The instrument will not allow sweep or list mode to be enabled at the same time that burst or any modulation
mode is enabled. When you enable sweep, the burst or modulation mode is turned off.
– To avoid multiple waveform changes, enable list mode after configuring its parameters.
Front Panel Operations
Enable list before setting any other list parameter. Press [Sweep] > Type
Linear
> Type List.
Keysight EDU33210 Series User's Guide
131