SCPI Command
[SOURce[1|2]:]FREQuency {<frequency>|MINimum|MAXimum|DEFault}
The APPLy command configures awaveform with one command.
Burst Count
Number of cycles (1 to 100,000,000 or infinite) to be output per burst. Used in the triggered burst mode only
(internal or external source).
– With the Immediate trigger source, the specified number of cycles are output continuously at a rate determined by
the burst period. The burst period is the time between the starts of consecutive bursts. Also, the burst count must be
less than the product of burst period and waveform frequency:
Burst Period > (Burst Count) / (Waveform Frequency) + 1 μsec
– The instrument will increase burst period to its maximum value to accommodate specified burst count (but
waveform frequency will not be changed).
– In gated burst mode, burst count is ignored. However, if you change the burst count from the remote interface
while in the gated mode, the instrument remembers the new count and will use it when the triggered mode is
selected.
Front Panel Operations
Press [Burst] > # of Cycles. Use the numeric keypad or the knob and arrows to set a desired value. If you use the
keypad, press Enter to finish.
Keysight EDU33210 Series User's Guide