2 Applies to sample clock, not whole waveform
– Frequency Limitations: Changing functions may change the frequency to meet the new function's frequency limits.
– Amplitude Limitations: When the output units are Vrms or dBm, changing functions may lower the amplitude to the
maximum for the new function due to variation in waveform shapes. For example, a 5 Vrms square wave (into 50 Ω)
changed to a sine will decrease to 3.536 Vrms (sine's upper limit).
– Amplitude and offset cannot combine to exceed the instrument's capability. The one you set last may be changed to
stay within limits.
– You may protect a device-under-test (DUT) by specifying upper and lower output voltage limits.
Front Panel Operations
– To turn on an output: Press Channel [On/Off] for your desired channel.
– To select another waveform: Press [Waveform].
For example, to specify a DC signal:
1. Press [Waveform] > MORE 1 / 2 > DC > Offset.
Use the numeric keypad or the knob and arrows to set a desired value. If you use the keypad, select a unit prefix to
finish.
Keysight EDU33210 Series User's Guide
81