SCPI Command
[SOURce[1|2]:]VOLTage {<amplitude>|MINimum|MAXimum|DEFault}
[SOURce[1|2]:]VOLTage:HIGH {<voltage>|MINimum|MAXimum|DEFault}
[SOURce[1|2]:]VOLTage:LOW {<voltage>|MINimum|MAXimum|DEFault}
The APPLy command configures a waveform with one command.
DC Offset Voltage
The default offset is 0 V for all functions.
– Limits Due to Amplitude: The relationship between offset voltage and output amplitude is shown below. The peak
output voltage (DC plus AC) cannot exceed the instrument output rating (±5 V into 50 Ω load, or ±10 V into an open
circuit).
– The relationship between offset voltage and output amplitude is shown below. Vmax is the maximum peak voltage
for the selected output termination (5 V for a 50 Ω load or 10 V for a high-impedance load).
|Voffset| < Vmax - Vpp/ 2
If the specified offset voltage is not valid, the instrument will adjust it to the maximum DC voltage allowed with the
specified amplitude. From the remote interface, a "Data out of range" error will also be generated.
– Limits Due to Output Termination: The offset range depends on the output termination setting. For example, if you
set offset to 100 mVDC and then change output termination from 50 Ω to "high impedance," the offset voltage
displayed on the front panel doubles to 200 mVDC (no error is generated). If you change from "high impedance" to
50 Ω, the displayed offset voltage will be halved. Changing the output termination setting does not change the
voltage present at the output terminals of the instrument. This only changes the displayed values on the front panel
and the values queried from the remote interface. The voltage present at the instrument's output depends on the
load connected to the instrument. See "OUTPut[1|2]:LOAD" in the EDU33210 Series Programming Guide for details.