GPIB Subsets
This instrument contains the following IEEE 488.1 subsets:
GPIB IEEE Std. 488.2 Error Handling – Query Error Register
The IEEE 488.2
If the instrument is addressed to talk and the response formatter is inactive and the input queue is
empty then the
the Standard Event Status Register, a value of 3 to be placed in the Query Error Register and the
parser to be reset. See the Status Reporting section for further information.
The IEEE 488.2
send a response message and a
or the input queue contains more than one END message then the instrument has been
INTERRUPTED
Standard Event Status Register, a value of 1 to be placed in the Query Error Register and the
response formatter to be reset thus clearing the output queue. The parser will then start parsing
the next
<PROGRAM MESSAGE UNIT>
further information.
The IEEE 488.2
a response message and the input queue becomes full then the instrument enters the
state and an error is generated. This will cause the Query Error bit to be set in the Standard Event
Status Register, a value of 2 to be placed in the Query Error Register and the response formatter
to be reset thus clearing the output queue. The parser will then start parsing the next
MESSAGE UNIT>
GPIB Parallel Poll
Complete parallel poll capabilities are offered on this instrument. The Parallel Poll Enable Register
is set to specify which bits in the Status Byte Register are to be used to form the
The Parallel Poll Enable Register is set by the *PRE <nrf> command and read by the *PRE?
command. The value in the Parallel Poll Enable Register is ANDed with the Status Byte Register; if
the result is zero then the value of
The instrument must also be configured so that the value of
during a parallel poll operation. The instrument is configured by the controller sending a Parallel
Poll Configure command (PPC) followed by a Parallel Poll Enable command (PPE). The bits in the
PPE command are shown below:
bit 7 =
bit 6 =
bit 5 =
bit 4 =
Source Handshake
Acceptor Handshake
Talker
Listener
Service Request
Remote Local
Parallel Poll
Device Clear
Device Trigger
Controller
Electrical Interface
error (addressed to talk with nothing to say) is handled as follows.
UNTERMINATED
error is generated. This will cause the Query Error bit to be set in
UNTERMINATED
error is handled as follows. If the response formatter is waiting to
INTERRUPTED
<PROGRAM MESSAGE TERMINATOR>
and an error is generated. This will cause the Query Error bit to be set in the
error is handled as follows. If the response formatter is waiting to send
DEADLOCK
from the input queue. See the Status Reporting section for further information.
X
don't care
1
1
Parallel poll enable
0
SH1
AH1
T6
L4
SR1
RL2
PP1
DC1
DT0
C0
E2
from the input queue. See the Status Reporting section for
is 0 otherwise the value of
ist
has been read by the parser
ist
is 1.
ist
can be returned to the controller
ist
DEADLOCK
<PROGRAM
local message
21