Troubleshooting
If the system stops working properly, check whether you are able to correct the
problem yourself based on the tips below. If this is not possible, contact the
customer service department.
Do not attempt to repair a electrical device yourself!
Problem
The cuffs do not
pump up within
15 minutes.
The cuffs do
not deflate
completely.
Possible cause
Air escapes between
the connection tubes
and tube connectors.
The pump is not
activated.
The air input is
obstructed.
The air extraction func­
tion is not activated.
The air flow is
obstructed.
Fix
Push the ends of the con­
nection tubes onto the tube
connectors of the cuffs and
control unit as far as they will
go.
Check to see if the ends of
the connection tubes are
firmly attached to the tube
connectors or whether they
have widened. If they have
widened, cut off the widened
ends.
Press the
button to pump up
the cuffs. Set the PRESSURE
regulator to the highest level.
Make sure the connection tubes
and cuffs are not bent or twisted.
Press the
button to extract air
from the cuffs. Set the PRESSU-
RE regulator to the highest level.
Make sure the connection tubes
and cuffs are not bent or twisted.
47