The actual endpoint format (last column) and not the preselected is stored with the data.
1
6.6  T imed Interval Readings
A reading is taken every time after a certain interval (1 – 2400 s) defined in the menu has
elapsed. When working in the timed-interval reading mode, the interval can be defined by entering
the seconds. The measurement series stops according to the selected endpoint format or manually
by pressing READ. When timed-interval reading is "on", the DL icon
The readings can be stored in the memory, transferred to the interface or both.
6.7  T emperature settings
Set MTC temperature
If the meter does not detect a temperature probe, MTC appears on the display. In this case the
sam p le temperature should be entered manually. An MTC value between -30 °C and 130 °C
can be entered.
Temperature unit
Select the temperature unit: °C or °F. The temperature value is automatically converted
between the two units.
The instrument works with NTC30 kΩ and Pt1000 temperature sensors. You can select be-
tween automatic recognition or manual selection of the temperature sensor type. For tem p er-
a t ures below 100°C the instrument can reliably distinguish between NTC30 kΩ and Pt1000.
At higher temperatures, however, it is necessary to set the type of temperature sensor man u a l-
ly to avoid mistakes.
6.8  M easurement limits
The upper and lower limits for measurement data can be defined. If a limit is either not reached or
exceed e d (in other words, less than or greater than a specific value), a warning is displayed on
the screen and may be accompanied by an acoustic signal. The message "outside limits" also ap-
pears on the GLP printout.
appears.
Menus and settings
33