Descargar Imprimir esta página

Lowenstein Medical Silentflow 3 Manual De Instrucciones página 26

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4.1 Cleaning the exhalation system
1. Dismantle the exhalation system
exhalation system", page
2. Clean the individual parts in accordance with the table below:
Action
Wash individual parts in hot water (min. 30 °C) and mild
detergent (1 ml detergent to 1 l water) for 15 minutes.
Clean individual parts thoroughly with a cloth or a soft brush
when washing.
Or: Put individual parts in the dishwasher (max. 65 °C).
3. Wash immersed individual parts with a soft brush for at least 3
minutes.
4. Pay attention to all creases and cavities.
5. Rinse the individual parts in clean water.
6. Allow the individual parts to air-dry.
7. Perform a visual inspection.
8. If necessary: Replace a damaged exhalation system.
9. Assemble the exhalation system
exhalation system", page
Discolorations do not impair the function of the exhalation
system.
|
26
EN
Hygiene treatment and disposal
(see "3.3 Dismantling the
25).
(see "3.1 Assembling the
23).
X
X

Publicidad

loading