Technical Specifications - Mettler Toledo AE160 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

TECHNICAL SPECIFICATIONS

Readability
Weighing range
Tare range (subtractive)
Stabilization time (typical)
Integration time (can be configured: 1/2/3)
Reproducibility (standard deviation)
Linearity
Linearity
relative to
Admissible ambient temperature during operation
Sensitivity drift
Stability detector (can be configured)
Power supply:
– Voltage, adjustable
– Admissible voltage range
– Frequency
– Power consumption
Weighing pan of stainless steel
Weighing chamber (width x depth x height)
Balance housing (width x depth x height)
Net weight
Calibration weight (built-in)
14
AE160
0.1 mg
0...162 g
0...162 g
5 sec.
1.5/3/6 sec.
0.1 mg
+ 0.2 mg
10 g
+ 0.1 mg
10...40 °C
10...30 °C
+ 2·10
2 sensitivity steps / switched off
115 V / 220 V
92...132 V / 184...265 V
50...60 Hz
approx. 10 VA
∅ 80 mm
172 x 155 x 220 mm
205 x 410 x 290 mm
10.3 kg
100 g, + 0.2 mg, adjusted to an apparent mass of 8.0 g/cm
AE100
0.1 mg
0...109 g
0...109 g
-6
/°C
density at an air density of 1200 mg/l.
3