Light Source - Troubleshooting - Storz 81 3010 21 Manual De Instrucciones

Fuente de luz led accionada por batería
Ocultar thumbs Ver también para 81 3010 21:
Tabla de contenido

Publicidad

Symptom
Light source cannot be connected to the MoVeo
videoscope
endoscope or flexoscope
use.
44
5.2
Light source – troubleshooting
Possible causes
An adaptor sleeve is screwed
onto the videoscope light port
pin
An incorrect adaptor sleeve is
screwed onto the light port pin
or there is no adaptor sleeve
for more than 30 minutes at
maximum light intensity
enough
Batteries either do not work or
not to a sufficient degree
Battery charge state is low
Remedy
Remove the screw-on adaptor
sleeve or, depending on the version,
use a suitable sleeve.
Reduce the light intensity if light
light source off for a few minutes to
allow the housing to cool down.
Replace the batteries with the
designated fully charged batteries.
Check that the batteries are
inserted correctly.
Check that the specified batteries
are used.
Check the battery charge state and
use fully charged batteries.
level.
Replace the batteries with the
designated fully charged batteries. If
to a lower level in order to continue
using the light source for a few
minutes and replace the batteries
with fully charged ones as soon as
possible.

Publicidad

Tabla de contenido
loading

Tabla de contenido