Electromagnetic immunity
During the immunity testing below the BEMER Therapy System Evo will continue to provide
stimulation normally.
Emission measurements
Discharge of static electricity in accordance
with IEC 61000-4-2
Electrical fast transient/burst immunity
in accordance with IEC 61000-4-4
Impulse voltage/surges in accordance
with IEC 61000-4-5
Magnetic field at supply frequency (50/60 Hz)
in accordance with IEC 61000-4-8
Voltage dips, short-term interruptions
and fluctuations in the supply voltage
in accordance with IEC 61000-4-11
Remark: U
is the AC mains voltage prior to applying the test level
T
Immunity tests
Conducted RF disturbance in accordance with
IEC 61000-4-6
Radiated RF disturbance in accordance with
IEC 61000-4-3
Radiated magnetic fields IEC 61000-4-39
Immunity to RF telecommunication equipment pursuant to limit values of IEC 61000-4-3 tested and passed.
2119-2208-764120_IFU Evo_USA_PEMF_764120-GC-3015_Inhalt_v08.indd 30
2119-2208-764120_IFU Evo_USA_PEMF_764120-GC-3015_Inhalt_v08.indd 30
Compliance level
± 8 kV contact discharge
± 2, 4, 8 and 15 kV air discharge
± 2 kV for power cables
± 1 kV voltage phase conductor –
phase conductor
30 A/m 50 and 60Hz
< 5 % U
for 1/2 period (> 95 % dip)
T
< 40 % U
for 10 periods (60 % dip)
T
< 70 % U
for 25 periods (30 % dip)
T
< 5 % U
5 s (> 95 % dip)
T
Compliance level
3 V
eff
150kHz to 80MHz inside the ISM/amateur band
6 V
eff
150kHz to 80MHz inside the ISM/amateur band
10V/m
80 Mhz to 2.7GHz
8 A/m at 30 kHz
65 A/m at 134.2 kHz
7.5 A/m at 13.56 kHz