INSTRUCTIONS FOR USE
5 PRE-INSTALLATION REQUIREMENTS
Caution
Pre-installation and installation must be performed according to the applicable standards of the particular
country and in accordance with the valid documentation of the manufacturer, which is owned by each
authorized representative of DIPLOMAT DENTAL s.r.o.
Caution
To avoid the risk of electric shock, device must be connected to the power supply with a protective earthing.
Do not modify this device without prior authorization of the manufacturer.
Environmental conditions
5.1
Not to be installed in the premises with explosion hazard!
Requirements for the installation of media
5.2
Water
Distilled water is required to be used.
Air
Supply of at least 55 l/min. of air at the pressure of 0,45 až 0,8 MPa, oilless, clean and dry must be ensured. Tubes
made of Cu and/or PE are recommended.
Recommended mains fuse rating
Recommended rating of the fuse of the supply main is 16A (in the event of circuit breaker – circuit breaker with
switching-off characteristic of C type). No other items of equipment should be connected to the supply main in
question!
Maximum power input of the dental unit is 150 VA. The supply main must comply with the respective national standard.
Recommendation
Unless the national standard stipulates otherwise, the manufacturer recommends to use current protective switch with
the sensitivity of 30mA and instantaneous time of switching-off.
Pre-installation requirement having been met, assembly and installation of the dental unit is carried out and it is connected
to the media.
Floor
5.3
The floor must have a concrete foundation of at least 100 mm thickness. The slope of the floor must not exceed 1%. The
use of antistatic floor is recommended.
Environment
5.4
ambient temperature range
relative humidity range
atmospheric pressure range
6 ASSEMBLY AND INSTALLATION
The installation must be done by the certified service technician only; otherwise no possible future warranty
claims will be accepted. The registration form must be filled out and sent to the manufacturer or the seller.
Unpacking of the unit and inspection of the delivery
Inspect the transport packages for damage. If a defect of the transport package is found, do not open the consignment
and report the defect to the forwarding agent or seller immediately. If the consignment is intact, carefully open the
package and unpack individual parts of the dental unit. Check the completness of the delivery according to the
Packing list.
UM_EN_TYGI350B_2018-09_ver1.0
from +10° C
from 30%
from 700hPa
TYGI 350 B
to +40° C
to 75 %
to 1060hPa
8 / 49