Descargar Imprimir esta página

Vaisala Indigo510 Guía Rápida página 10

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 91
Property
Selectable current output types
Max. wire size
Accuracy of analog outputs at +20 °C (+68 °F)
Temperature dependence
External loads:
Current outputs
0 ... 1 V output
0 ... 5 V and 0 ... 10 V outputs
Indigo520 relay outputs
Number and type of relays
Max. switching power, current, voltage
Max. wire size in PELV version
Max. wire size in AC (mains) version
Ethernet interface
Supported standards
Connector
Supported protocols
1)
The power supply option is selected when ordering the transmitter.
Table 3 Compliance
Property
Indigo520 EU directives
Indigo510 EU directives
EMC compatibility
Electrical safety
Indigo520 compliance marks
Indigo510 compliance marks
Listing marks
8
Description/Value
4 ... 20 mA, 0 ... 20 mA, scalable
2
2.5 mm
(14 AWG)
±0.05 % full scale
±0.005 % / °C full scale
R
< 500 Ω
L
R
> 2 kΩ
L
R
> 10 kΩ
L
2 pcs, SPDT
30 W, 1 A, 40 V DC / 28 V AC
2
2.5 mm
(14 AWG)
2
1.5 mm
(16 AWG)
10BASE-T, 100BASE-TX
8P8C (RJ45)
Modbus TCP/IP (port 502), HTTPS (port 8443)
Description/Value
EMC Directive (2014/30/EU)
Low Voltage Directive (2014/35/EU)
RoHS Directive (2011/65/EU) amended by
2015/863
EMC Directive (2014/30/EU)
RoHS Directive (2011/65/EU) amended by
2015/863
IEC/EN 61326-1, industrial environment
CISPR 32 / EN 55032, Class B
IEC/EN 61010-1
CE, China RoHS, FCC, RCM
CE, China RoHS, FCC, RCM, UKCA
Canada (SGS), US (SGS)
M212290EN-D

Publicidad

loading

Este manual también es adecuado para:

Indigo500 serieIndigo520