Use of the microscope in phase contrast
Installing the phase contrast slider
1.
Insert the slider into the lower slot of the illumination sys-
tem, printed face up. (Fig. 24)
2. Pull the slider into the desired position, until it arrives to the
click stop.
3. When in phase contrast observation, keep the aperture di-
aphragm adjustment lever
Phase contrast slider
•
The phase ring is pre-centered when the microscope leaves
the factory. It should therefore need no further adjustment.
Should a re-centering is needed, it can be performed via
the two side bolts (see Chapter "Centering the phase ring").
The 4x/10x position ② must be used with 4x and 10x phase
•
contrast objectives, the 20x/40x position ③ with the 20x
and 40x and the SL position ④ is used for brightfield. (Fig.
25)
SLIDER
MEANING
POSITION
SL
empty hole
4x/10x
phase ring 4x/10x
20x/40x
phase ring 20x/40x
Centering the phase ring
•
Usually this operation is not needed. If necessary, please
proceed with the following steps:
1.
Place a specimen on the stage and focus it.
2. Take out the eyepiece from the eyepiece tube without the
diopter adjustment, and replace it with the centering tele-
scope (CT). (Fig. 26)
3. Check that the phase ring and the objective match, and
that both are steadily set on a click stop.
vwr.com I Instruction manual INVERTED MICROSCOPE, VISISCOPE
①
on the "O" (open) position.
APPLICATION
brightfield
observation
phase contrast
observation with 4x
and 10x objectives
phase contrast
observation with 20x
and 40x objectives
①
②
IT417 PH
®
F ig. 24
F
ig. 24
③
④
F ig. 25
F
ig. 25
F ig. 26
F
ig. 26
20