Domus 2+ / English
7. Maintenance
General
7.1.
Check main power cord and plug if there is abrasions or excessive wear.
(1)
Check mattress cover for signs of wear or damage. Ensure mattress cover and
(2)
tubes are stubbed together correctly.
Check airflow from the quick connector. The airflow should alternate between
(3)
each connector every half-cycle time if it's in alternating mode.
Check the air hoses if there is kink or breaks. For replacement, please contact
(4)
our local dealers.
Fuse Replacement
7.2.
Air Filter Replacement
7.3.
Disconnect the plug form mains power when a
(1)
fuse is to be replaced.
Remove the cover of the fuse holder by means
(2)
of a screwdriver. (If necessary, un-clip the
hanger form the pump unit)
Insert a new fuse of the correct rating in, and
(3)
replace the cover of the fuse holder back. The
fuse should be T1AL/250V type and VDE
approved.
Open the air filter plate located at the back of
(1)
pump.
Suggest to check and replace air filter
(2)
regularly
if environment is dirty.
(3)
11