Descargar Imprimir esta página

Wellell Domus 2+ Manual De Instrucciones página 18

Sistema de redistribución de presión alterna

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
Domus 2+ / English
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user
of this device should make sure it is used in such an environment.
Immunity Test Levels
Professional
Basic EMC
healthcare
standard
facility environ-
ment
Electrostatic
Discharge
±8kV contact
(ESD)
±15kV air
IEC61000-4-2
Electrical
fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to
earth
Voltage
dips,
Voltage Dips:
short
interrup-
i) 100% reduction for 0.5 period,
tions and volt-
ii) 100% reduction for 1 period,
age
variations
iii) 30% reduction for 25/30 period,
on power sup-
Voltage Interruptions:
ply input lines
100% reduction for 250/300 period
IEC61000-4-11
Power
fre-
quency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM
Conducted RF
bands
IEC 61000-4-6
between 0,15
MHz and 80
MHz
80 % AM at 1
kHz
Compli-
HOME
ance
HEALTHCARE
Levels
ENVIRONMENT
±8kV
contact
±15kV
air
±2kV for
power
supply
line
±1kV for
in-
put/out-
put line
± 1 kV
± 1 kV line(s) to
line(s) to
line(s)
line(s)
230V
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and
amateur
radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Electromagnetic Environment-
Guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic ma-
terial, the relative humidity
should be at least 30 %.
Mains power quality should be
that of atypical commercial or
hospital environment
Mains power quality should be
that of atypical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of this device requires
continued operation during
power mains interruptions, it is
recommended that the device
be powered from an uninter-
ruptible power supply or a bat-
tery.
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location
in a typical commercial or hos-
pital environment.
Portable and mobile RF com-
munications equipment should
be used no closer to any part
of this device, including ca-
bles, than there commended
separation distance calculated
from the equation applicable
to the frequency of the trans-
mitter.
15