of water bad quality from the autoclave, the filter‐cartridge assembly has to 
be replaced as follows: 
 
 Switch off the machine and close the cock on the tank. 
 Remove  the  safety  clips  from  the  inlet  and  outlet  couplers  of  the 
exhausted filter‐cartridge assembly   
 Disconnect  hoses  connecting  the  water  line  dedicated  to  the    filter‐
cartridge assembly  by pressing to the coupler body the stop ring and 
then pulling the hose. 
 Replace the exhausted filter‐cartridge assembly  with a new one. 
 Reconnect the hoses of the water line to the filter‐cartridge assembly 
by fully fitting them. 
 Apply safety clips to the quick couplers. 
 Temporarily  disconnect  the  filter  from  the  resins  cartridge  and  take 
the filter free end on a basin or a big container 
 Re‐open the cock on the tank and restart the Machine. 
 Run  water  through  the  filter  until  water  coming  out  is  completely 
limpid. 
 Switch off the Machine and close the cock on the tank. 
 Re‐connect  the  filter  to  the  cartridge  by  fully  pressing  the  hose  and 
fitting the safety clip 
 Re‐open the cock on the tank and restart the Machine. 
 Be sure of the absence of water leaks through couplers. 
 Disassemble the exhausted assembly by separating the GAC filter from 
the resins cartridge and dispose of the two exhausted parts following 
indications of  Tab. 8.1.1 of this Manual.
 