of water bad quality from the autoclave, the filter‐cartridge assembly has to
be replaced as follows:
Switch off the machine and close the cock on the tank.
Remove the safety clips from the inlet and outlet couplers of the
exhausted filter‐cartridge assembly
Disconnect hoses connecting the water line dedicated to the filter‐
cartridge assembly by pressing to the coupler body the stop ring and
then pulling the hose.
Replace the exhausted filter‐cartridge assembly with a new one.
Reconnect the hoses of the water line to the filter‐cartridge assembly
by fully fitting them.
Apply safety clips to the quick couplers.
Temporarily disconnect the filter from the resins cartridge and take
the filter free end on a basin or a big container
Re‐open the cock on the tank and restart the Machine.
Run water through the filter until water coming out is completely
limpid.
Switch off the Machine and close the cock on the tank.
Re‐connect the filter to the cartridge by fully pressing the hose and
fitting the safety clip
Re‐open the cock on the tank and restart the Machine.
Be sure of the absence of water leaks through couplers.
Disassemble the exhausted assembly by separating the GAC filter from
the resins cartridge and dispose of the two exhausted parts following
indications of Tab. 8.1.1 of this Manual.
49
ENGLISH