GENERAL INFORMATION 
 
Fitting to be supplied by the customer :  
 
 
Unless differently agreed upon by contract, the Customer shall provide for: 
 
‐ Preparation of the premises, including any brickworks and/or canalization required. 
 
‐ Power supply of the machine in compliance with the rules in force in the use country: 
the machine is fitted with two type L 10 A plugs (3‐pole Italian plug), complying with 
the  CEI  23‐16/VII,  10°  rules.  Any  connector  for  different  types  of  plugs  shall  be 
provided for by the Customer, in compliance with rules in force in the use country . 
 
‐ Water feeding, water discharge system and distribution of the treated water to the 
utilities: in this aim a water system satisfying the following characteristics and 
instructions shall be provided: 
‐ Next to the machine installation area a water usage point from the water distribution 
network has to be prepared with closing device (usually a ball valve) fitted with a ¾" 
external thread coupling, like those commonly used for the connection of washing 
machines, dish washers, or cleaners. The water flow at the usage point must not be 
lower than 300 l/h, otherwise serious machine malfunction may occur.
‐ Next to this point there will be the infeed of the water network destined to supply 
treated water  to the dental units and, if any, to an autoclave and/or an additional cock 
on the basin located in the sterilization room. This usage point too shall be fitted with 
ball valve and a  3/8" or ½" internal thread coupling.
‐ A by‐pass cock is necessary to connect the water distribution network upstream the 
usage point and the associated valve with the network downstream the usage point of 
the treated water and associated valve, to allow the use of dental units with the 
network distribution system in case the Machine is out of order due to repair or 
maintenance interventions. 
‐ Next to the couplings to the water distribution network a collection point for waste 
water from the Machine has to be provided having a discharge capacity of 300 l/h of 
water without remarkable counter‐pressures. 
 
‐ The water line dedicated to the supply of dental units and , if any, to the autoclave 
and/or additional cock of the basin in the sterilization room shall have to be made in 
food‐grade plastic material, as well as the relevant fittings. Iron, copper, or brass 
fittings are not allowed. For fittings only , in case plastic fittings are not available, use 
AISI 316 stainless steel or chromium plated brass (less advisable). 