Starting and testing (Fig. 2):
After completing the hydraulic and electric connections follow instructions below 
 Do not switch On the Biox 
 Position  the  Opening  Lever  of  the  tank  discharge  [14]  to  "Open" 
(horizontal position). 
 Close the cock on the tank. 
 Close the supply cock to the practice equipment. 
 Open the by‐pass cock. 
 Open the cock for main body supply 
 Power on. 
 Be sure the light of the discharge solenoid valve [4] is on; otherwise start 
the solenoid valve switch [5] to light it up. 
The  pump  inside  the  main  body  runs  for  some  seconds  then  stops.  The 
pressure  gauge  on  the  front  panel  of  the  main  body  indicates  the  pressure 
 
value for some seconds. 
 Now open the cock on the tank only. 
The  pump  restarts  and  production  of  purified  water  starts  until  the  tank  is 
full, then the system stops . 
 Now  close  the  by‐pass  cock  and  open  the  supply  cock  to  the  supplied 
devices. 
Pick  water  through  the  supplied  devices  or  the  cock  on  the  basin  until 
causing the pump restart which will stop as soon as the set water reserve 
is refilled. 
 
The machine is now ready for use. 
 
 
 