Error message
Err 5
bAtt
Err 6 +
possible
continuous alarm
until a button is
pressed
Possible cause
1. Power pack or battery voltage
too low
2. Power packs or batteries
defective
3. Battery contacts are corroded
1. Build-up of air
2. Blood pressure cuff incorrectly
connected
3. Leak in the cuff or connecting
tube
Troubleshooting
Measures
1. Replace the power packs or
batteries
2. The power pack or battery
voltage is correct but bAtt is shown
on the display during cuff inflation.
Replace the power packs
3. Clean the battery contacts with a
cotton cloth and a little alcohol.
1. Check the cuff for a build-up of
air or a kink in the tube. If there is a
kink in the cuff tube, lay the tube
out. Otherwise, send the device in
immediately.
2. Connect the cuff to the device
(see chapter 3.3.5 "Cuff
connection")
3. If necessary, replace the cuff or
connecting tube.