RECOMMENDED MAINTANCE
In the interest of equipment safety, a maintencnce program must be established. It is the
owner's responsibility to arrange for this service and to assure that the personnel
performing this are fully qualified to service GENDEX dental x-ray equipment.
The calibration and adjustments to be performed are listed in the System and Function
Checks section of the 765DC Installation and Maintenance manual supplied with the
product.
It is recommended that the following System and Function Checks and
Maintenance Procedure be performed at installation and every twelve months
thereafter.
System Function Checks listed in Section E, page 35 of the 765DC
Installation / Maintence Manual.
Suspension Arm Maintenance as listed in Section F, page 43 of the 765DC
Installation / Maintenance Manual.
Note: To avoid any potential hazard to operators or patients, any unusual
operation, mechanical functionality, or debris should be reported to your service
agent immediately.

CLEANING AND DISINFECTING

All external and cleanable surfaces can be cleaned and/or disinfected using the
following list of commonly used solutions:
Isopropyl alcohol
Phenylphenl
Benzyl-P-Chlorophenol
Iodophor
Disobutyl phenoxethoxycthyl Dimethyl Benzyl Ammonium Chloride
Note: Use bleach or glutaraldehydes only occasionally.
Verify the contents of any cleaning or disinfecting solution against the list
above applying it the any surfaces of the 765DC.