Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The No Touch is intended for use in the electromagnetic environment specified below.
The customer or the user of the No Touch should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site sur-
vey should be considered. If the measured field strength in the location in which the No Touch is used exceeds the
applicable RF compliance level above, the No Touch should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the No Touch.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
100022_MEBBY NO TOUCH USER MANUAL_REV.00_21.04.2016.indd 131
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should
be used no closer to any part of the No Touch, including
cables, than the recommended separation distance calcu-
lated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Not applicable
d = 1.2
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmit-
3 V/m
ter in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey, a should be less than the
compliance level in each frequency range b.
Interference may occur in the vicinity of equipment
marked with the following symbol:
131