Idiomas disponibles

and press [Print] to decrease the value. The internal calibration weight should
only change between 95.000 and 105.000 grams or 495.00 and 505.00 grams
(depending upon the model).
6) Press [Tare] to store a particular setting. The display will return to F6
CA
7) When F6 CA is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
Changing the value of the internal weight will affect the calibration when using the
internal calibration. To check if your internal weight is correct you should only use
high quality weights that are accurate to the readability of the balance.
6.7

SETTING USER PARAMETERS

The balance has a number of parameters the user can adjust.

Display Description

Filter settings, 1, 2 or 3
n FIL
1 is the fastest filter, 3 is the slowest.
The Automatic Rezero range
n ZEO
0 to 8, 0 being with auto zero off, and 8 is the zero
range of 4 divisions.
Stability symbol range, 0 to 5
n STA
0 is the smallest stability range and 5 is largest.
n ZTR
Weight Tracking Range, 0 to 5
Press the [Tare] key to view the current settings.
To change the settings press the [Mode] key.
the next parameter. When completed the display will show F7 SET again.
© Adam Equipment Company 2014
Press the [Tare] key to advance to
19
9520revF-Jun14
Default Value
1 FIL
4 ZEO
2 STA
3 ZTR