Maintenance; Manutenzione / Wartung; Maintenance Interval; Maintenance Procedure - Ralston Instruments DP0V Serie Manual De Operación

Bombas manuales de cilindro neumático
Tabla de contenido

Publicidad

Maintenance

Maintenance ....................................17
Mantenimiento .................................18
維護 .................................................18
Техническое обслуживание ..........18

Maintenance Interval

EN
Every 300 uses or 3 months

Maintenance Procedure

• Lubricate External Check Valve by squirting 2 ml of oil into it, and then
drawing the oil into the Check Valve by pumping the piston several times.
• Lubricate Internal Check Valve by removing pumping piston and squirting
2 ml of oil into hole in Check Valve at bottom of pumping piston cavity.
When pump is pumped oil will be drawn into the Check Valve.
• Lubricate the Ralston Quick-test™ fittings by squirting 2 ml of oil inside
the connection.
• Lubricate the pumping piston and the fine adjust piston O-rings with
silicone lubricant.

Intervalle d'entretien

FR
Tous les 300 usages ou 3 mois
Procédure d'entretien
• Lubrifiez le clapet antiretour externe en déposant 2 ml d'huile dedans,
puis en tirant l'huile dans le clapet antiretour en pompant le piston
plusieurs fois.
• Lubrifiez le clapet antiretour interne en retirant le piston de pompage et en
déposant 2 ml d'huile dans le trou du clapet antiretour en bas de la cavité
du piston de pompage. Lorsque la pompe sera activée, l'huile sera tirée
dans le clapet antiretour.
• Lubrifiez les raccords Ralston Quick-test™ en déposant 2 ml d'huile à
l'intérieur de la connexion.
• Lubrifiez les joints toriques du piston de pompage et du piston de réglage
micrométrique avec du lubrifiant à la silicone.
Manutenção .....................................19
Manutenzione ..................................19
Wartung ...........................................20
17

Publicidad

Tabla de contenido
loading

Tabla de contenido