Technical Specifications - TENTECH MPK-254 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

10. Technical specifications

Test currents
Resistance ranges
Resolution
Output voltage
Measurement principle
Basic accuracy
Advanced features
Serial data output
Environmental protection
Safety class
Power supply
Built-in battery charger
Operating temperature
range
Storage temperature range : -25° C to 65° C.
Humidity range
Weight
: 1 mA - 10 mA - 100 mA - 1 A - 5 A.
Each current may be continuously adjustable from
0 to 100%.
: 0 - 20 m
@ 5 A.
0 - 600 m
@ 5 A.
0 - 200 m
@ 1 A.
0 - 2000 m
@ 100 mA.
0 - 20
@ 10 mA.
0 - 200
@ 1 mA.
: 1 µ @ 5 A.
: Up to 10 Vd.c. @ 1 A (open circuit).
: Four-terminal, Kelvin-type.
: ± 0.2% of reading ± 2 digits.
: Digital direct reading of very low resistances in the
alphanumerical display, with up to 4½ digits. Very
fast and accurate measurements.
: RS232 @ 4800 bps. Suitable for data collection in
an external serial printer, computer or data-logger.
: IP54 (with closed lid).
: Meets the requirements of IEC 61010-1.
: Rechargeable battery 12 V - 7 Ah.
: 220-240 V~ mains supply.
: -5° C to 50° C.
: 95% RH (non condensing).
: Approx. 8 .7 kg (including accessories).
14

Publicidad

Capítulos

Tabla de contenido
loading