Technical Data - IMT Analytics SmartLung 2000 2L Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Preface
The SmartLung 2000, SmartLung Adult and the Smart-
Lung Infant are the same size as conventional test lungs
but offer both an adjustable compliance and a resistan-
ce that can be varied between Rp5 and Rp200. The
adjustable leakage also enables the user to verify the
leak compensation of the ventilator. The patient trigger
functions can also be verified.
The SmartLung is the ideal tool for verifying ventilators
and anaesthesia equipment and is the ideal supple-
ment for CITREX and FlowAnalyser flow and pressure
measurement devices.
Typical settings
Adult
Compliance
(mL/mbar)
Compliance
(mL/kPa)
Resistance
Paediatric
Compliance
(mL/mbar)
Resistance
Infant
Compliance
(mL/mbar)
Resistance

Technical Data

Compliance
(mL/mbar)
Resistance
Volume (mL)
Leak (L/min)
Weight (g)
Dimensions (mm) 400×140×40 303×115×40 270×115×40
Deadspace (mL)
Components
Cleaning
* Tolerances: Resistance ± 10 %, compliance ± 20 %
SmartLung
SmartLung
2000 2 L
Adult 1 L
25, 60, 75
10, 15, 20,
30
200, 500
Rp5, Rp20,
Rp5, Rp20,
Rp50
Rp50
10, 15, 20
Rp20, Rp50
SmartLung
SmartLung
2000 2 L
Adult 1 L
25, 60, 75
10, 15, 20,
30
Rp5, Rp20, Rp50, Rp200
0 –1000
0 – 600
2 L bag
1 L bag
400
325
replaceable
by disinfection
SmartLung
Infant 0.5 L
1, 2, 3, 5
Rp50,
Rp200
SmartLung
Infant 0.5 L
1, 2, 3, 5
0 – 200 mL
0.5 L bag
0 – 10
285
3.7
Page 17